食品飲料品質控制檢驗儀器 系列 Quality Control Instruments for Beer & Beverage 
 
 

 

梅瑟利 Maselli Misure 製程線上及實驗研究品質控制之自動監測糖度儀 / 折光儀 / 濃度儀
PROCESS ANALYZERS Instruments and Systems for Liquid Analysis

 

 

 

BEER ANALYSIS SYSTEMS / 製程線上多功能啤酒自動分析儀器系統
beer monitor IB04 analysis unit

The Maselli IB04, installed between the filler and the storage tank, achieves the analysis of all the fiscal parameters needed by the operator combing the measurement provided by a sound velocity transducer, a
refractometer, a carbometer and a O
2 electrode.
The unit can perform analysis of Alcohol, Extract, Original Gravity, CO
2 and O2 with alarm and data transmission functions.

原廠規格說明書下載 (PDF 185 KB)
原廠啤酒製程技術訊息資料下載 (PDF 310 KB)
beer monitor IB04 analysis unit
儀器技術規格 / Technical Specifications:
Measurement limits:
Alcohol:0...10% w/w; 0...12% V/V.
Extract: 0...10 Plato.
Original Gravity :0...30 Plato.
Carbonation: 0...5V/V; 0...10g/l.
Oxygen: 0...150 ppb.
Accuracy:
Alcohol: better than +/-0.02% w/w
Extract: better than +/-0.02 Plato.
Original Gravity: better than +/-0.05 Plato.
Carbonation: +/-0.05 V/V; +/-0.1g/l.
Oxygen: better than +/-2 ppb
Measurement scales:
Alcohol; PLATO; Extract;
"Gas/Vol" or "g/l";
O
2 ppb.
Product temperature:
-5°C...+35°C equivalent to 23°F...95°F with automatic compensation of
sugar concentration by means of a Pt100 temperature probe.
Line Pressure:
0...7bar (0...110psi) at 20°C (68°F).
Protection Category:
IP55 to EN60529.
Conformity:
CE mark shows conformity to listed UE Directives and standards.
Power Supply:
AC 3/PE 230/400V +/-10%, 50Hz+/-2% - 460V+/-10% 60Hz+/-2%, 1200VA
(other power supplies on request).
Interfaces:
N°2 4...20mA on 470 completely independent channels
RS422/485 for connection to the "CM00" remote Control Unit
RS422 interface "Labtech" data acquisition Software Package to a PC
RS485 interface (Optomux Protocol)
Characteristics: - Measurement prism in "Optical Glass".
- Light source (LED); High resolution detecting element "CCD".
- Backlit graphic LCD display a 240x128 points.
- "Mechanical Shock" degassing system.
- Relative Pressure Sensor in AISI 316 stainless steel & Hastelloy
C276.
- Self-priming centrifugal pump.
Dimensions:
660 (b) x 1650 (h) x 550 (p), ~150kg.
 
 
beer monitor basic solution

The analysis of Alcohol, Extract and Original Gravity is performed by the microprocessored controller of the system, able to combine the measurements coming from a sound velocity transducer and a refractometer.
The installation of the monitor is realized directly in line in the main piping using a standard Varivent fitting.

Optional additional parameters: CO
2 and O2.

 

UR20+US01 beer monitor basic solution

儀器技術規格 / Technical Specifications:
Measurement limits:

Alcohol:0...10% w/w; 0...12% V/V.
Extract: 0...10 Plato.
Original Gravity :0...30 Plato.

Accuracy:
Alcohol: better than +/-0.02% w/w
Extract: better than +/-0.02 Plato.
Original Gravity: better than +/-0.05 Plato.
Measurement scales:
Alcohol; PLATO; Extract.
Product temperature:
-5°C...+35°C equivalent to 23°F...95°F with automatic compensation of
sugar concentration by means of a Pt100 temperature probe.
Line pressure:
1...10bar (14,5...145psi) at 20°C
1...8bar (14,5...116psi) at 100°C.
Protection Category:
IP65 to EN60529.
Conformity:
CE mark shows conformity to listed UE Directives and standards.
Power Supply:
AC 18V (12...24V) 50...60Hz, 0.6A.
DC 24V (18...36V) 0.6A.
AC 115/230V +/-10% 50...60Hz, 10VA (Optional)
Interfaces: - 0...20mA or 4...20mA
- RS422/485 (Optomux Protocol)
- N°1 "Line Stop" input (HOLD); N°1 relay output for alarm signal.
Characteristics: - Measurement prism in "Optical Glass".
- Light source (LED); High resolution detecting element "CCD".
- Backlit graphic LCD display a 128x24 point.
 
 
CM00 Remote control and data acquisition software

製程線上多功能啤酒自動分析儀器系統之數據資料處理控制軟體

The newest and most complete software for remote control and data acquisition available on the market.

 

CM00 Remote control and data acquisition software

儀器技術規格 / Technical Specifications:
Conformity:
CE mark shows conformity to EU Directives
Supplies:
AC 115/230V+/- 10%, 50...60Hz, 600VA
Interfaces:

- RS232 or RS422/485 for connection to data acquisition Personal Computer
- RS422 for connection to "IB02" Beverage Analysis Unit and "QR20" Beverage Recording Unit
- Parallel CENTRONICS for connection to Printer and "Hardware key"
- AUX for Modem connection with input for auxiliary telephony device
- N. 8 relay outputs for signaling alarm completely configurable by means of software

Features: - Desktop Personal Computer, "CPU" with PENTIUM® Microprocessor, 16Mb Ram, Video Board, Modem, 3-button Mouse
- 17" Digital Monitor
- Hardware Key for Software protection
- Alarms output Interface Card
- Operating Software in Microsoft® Windows ambient
- Complete Monitoring and Management of a maximum of 8 Analyses and Regulation lines
- Creation and storage of a maximum of 999 combinations of operating parameters
- Tele-diagnostics or remote control of all functions by means of Modem+ Pcanywere software supplied