URX1 INTRINSICALLY SAFE REFRACTOMETERS / 防爆型製程線上自動連續監測折光計 | ||||
|
||||
儀器技術規格 / Technical Specifications: |
Measurement limits: |
-1.3330... 1.5318nD (0...95Brix).
|
Scale
Range Span: |
Min. 0.0227nD (15Brix)
Max 0.1988nD (95Brix). |
Accuracy: |
0.3% of Scale Range; maximum accuracy +/-0.00007nD +/-0.05Brix.
|
Measurement scales: |
"nD", "BRIX" and "USER"; "BRIX" scale refers to the nD/Bx ICUMSA (1974)
conversion tables "USER" scale can be configured at the time of order. |
Product
temperature: |
temperature measured by Pt1000 Temperature Sensor. |
Line
Pressure: |
1...8bar (14,5...116psi) at 100°C up to 25 bar upon request |
Protection
Category:
|
IP65 to
EN60529.
|
Conformity: |
CE mark
shows conformity to listed UE Directives and standards.
ATEX II 1 G - EEx ia IIB T4 |
Power
Supply: |
Using two 24 V galvanic separation
barriers connected in parallel mod. GM D1043Q or equivalent |
Interfaces: |
- Analog
(optional): 4-20 mA (max. load 470 Ohm)
with 3-wire galvanic separation barrier mod. GM D1010D or equivalent - Digital (optional): RS485 signal from the URX1 (in the hazardous area) is directly converted (in the safe zone) to RS232 or RS422/485 by the barrier. A module for the PROFIBUS DP interface, per EN5017, can also be connected to the barrier itself. Powered with galvanic separation barrier mod. GM D1043Q or equivalent. Signals using galvanic separation barrier mod. GM D1061S or Equivalent. |
Characteristics: |
-
Measurement prism in "Synthetic Sapphire" or "Optical Glass".
- Light source (LED); High resolution detecting element "CCD". - Local graphic LCD at 128x64 point. |