Measurement limits: |
1.3330... 1.5318nD (0...95Brix). |
Scale Range Span: |
Min. 0.0482nD (30Brix) - Max 0.1988nD (95Brix) "Standard Version".
Min. 0.0227nD (15Brix) - Max 0.1988nD (95Brix) "High Accuracy
Version". |
Accuracy: |
0.5% of Scale Range; maximum accuracy +/-0.0002nD +/-0.15Brix
"Standard Version".
0.3% of Scale Range; maximum accuracy +/-0.00007nD +/-0.05Brix "High
Accuracy Version". |
Measurement scales:
|
"BRIX" or "USER"; "BRIX" scale refers to the nD/Bx ICUMSA (1974)
conversion tables "USER" scale can be configured at the time of
order. |
Product temperature: |
-5°C...+105°C (140°C on request) with automatic compensation of
temperature measured by Pt100 Temperature Sensor. |
Line Pressure:
|
1...10bar (14,5...145psi) at 20°C
1...8bar (14,5...116psi) at 100°C. |
Protection Category: |
IP65 to EN60529. |
Conformity: |
CE mark shows conformity to listed UE Directives and standards.
"3A" USDA approval (on request). |
Power Supply: |
AC 18V (12...24V) 50...60Hz, 0.6A.
DC 24V (18...36V) 0.6A.
AC 115/230V +/-10% 50...60Hz, 10VA (Optional) |
Interfaces: |
- 0...20mA or 4...20mA
- RS422/485 (Optomux Protocol)
- N°1 "Line Stop" input (HOLD); N°1 relay output for alarm signal. |
Characteristics: |
- Measurement prism in "Synthetic Sapphire" or "Optical Glass".
- Light source (LED); High resolution detecting element "CCD".
- Backlit graphic LCD display a 128x24 point. |
Standard version: |
ø165 (b) x 180 (275) (h) x 130 (p), ~3.3kg. |
Tank Version: |