SENSOR
Velocity Measurement
Method: Electromagnetic (Faraday’s Law)
Range: -5 to +20 feet per second (-1.5 to
+6.1m/s)
Accuracy: ±2% of reading, ± 0.05 zero
stability at 0 to
+10 ft/sec (0 to + 3m/sec)
Zero Stability: ±0.05 feet per second (±0.015m/sec)
Resolution: 0.01 feet per second (±
0.003m/sec)
Level Measurement
Method: Submerged pressure transducer
Range: Standard 0.4 to 138 inches. (10mm
to 3.5m)
Optional 0.4 to 276 inches
Accuracy: ±1% of reading
Zero Stability: ± 0.03 ft. (± 0.009m)
zero stability for
0 to 10 ft (0 to 3m). Includes non-linearity,
hysteresis
and velocity effects.
Resolution: 0.1 inch (2.5mm)
Over Range Protection: 2 X range
Flow Measurement
Method: Conversion of water level and
pipe size to fluid
area. Conversion of local velocity reading to mean
velocity. Multiplication of fluid area by mean velocity
to equal flow
rate.
Conversion Accuracy: ±5.0% of reading.
Assumes
pipe is 10 to 90% full and with a level greater than
2 inches
Temperature Measurement:
Method: 1 wire digital thermometer
Range: 14º to 185ºF (-10º
to 85º C)
Accuracy: ± 6 F (3.5C) degrees
Material
Standard: Polyurethane
Dimensions: 5 3/16"(131.8mm) L, 1
¾ " (44.5mm)W,
1 1/8" H (29mm)
Weight: 2.5 lbs. (with 30 feet of sensor
cable)
Operating Temperature Range: 32° to 113°
F
(0° to 45° C)
Storage Temperature: -4º to 125º
F (-20° to 52°C)
Sensor Cable
Material: Polyurethane jacketed
Standard Length: 30 ft. (9m)
Optional Length: 60 ft. (18m),100 ft.
(30m) or length
as needed (maximum 1000 ft/300m)
Flow measurement is based on the equation:
Q = V x A where
Q = Flow
V = Average Velocity
A = Area.
The measurement of both velocity and level in the same
cross-section of pipe is critical for accurate flow rate determination,
including surcharge conditions.
|
|
FLO- LOGGER
Data Storage
64K (16K cycles of velocity/level data)
Local Terminal
RS232C at 19.2K baud
Power Requirements
Two 6-volt lantern batteries
Battery Life
Typical - 16 weeks at a sample interval
of 15 minutes
(with recommended batteries)
Housing
Sealed watertight PVC – meets IP68
rating
Length: 13.85 in. (35.179 cm)
Diameter: 7.75 in. (19.685 cm)
Weight: 7.5 lbs. (3.4 kg) (including
batteries)
Temperature
Operating Range: 14° F to 125° F (-10°
C to 51° C)
Storage Temperature Range: -4° F to 125°
F
(-20° C to 51° C)
Sensor/Logger Disconnect
Both the sensor and the logger have
waterproof (IP68)
connectors for easy separation from the
interconnecting
cable.
Certification
FCC Rules Part 15
Industry Canada RSS 210
No user license required
FLO-STATION
Data Storage
64K (16K cycles of velocity/level data)
Local Terminal
RS232C at 19.2K baud
Outputs
(4) 4-20 mA isolated outputs scaleable to
desired values
and capable of sinking 50 ohms each. Contact Closure
(One
form C dry).
Power Requirements
+12VDC @220 mA with three 4-20mA outputs
@ 20Ma
Housing
NEMA 4X (IP66/67-7)
Material: Polystyrene
Dimensions: 9 in. W x 7 in. H x 4 in. D
(229 mm W x 178 mm H x 102 mm D)
Weight: 2.75 lbs. (1.25 kg)
Temperature
Operating Temperature Range: 14° to 122°F
(-10° C to 50° C)
Storage Temperature Range: -4° to 122°F
(20° C to 50° C)

|