HygroGuard
2650
Trace
Moisture Analyzer
The HygroGuard 2550 is a
microprocessor-controlled trace moisture analyzer and a remote probe
which houses a gold/aluminum oxide capacitive sensor and a Zener diod
temperature sensor. The HygroGuard 2550 was designed to provide
economical trace moisture analysis for gases. Output range, alarm set
points, measurement units (dew point or ppmv) and temperature
calibration data are all user-programmable.
The instrument measure dew point temperatures from -80°C to +20°C
(-112°F to +68°F), with special calibration available to -100°C (-148°F),
with ± 2°C (± 3.6°F) accuracy. Process control is obtained with one
alarm relay and one, isolated, linear 4-20 mA output corresponding to
dew point or ppmv. A second alarm is also assignable to the m oisture
unit. Three push-butons allow the user to easily set the alarm set
points, measurement unit, analog output range, temperature calibration
data, and pressure constant. The instrument electronics are EMI/RFI/ESD
resistant, exceeding IEC and ANSI/IEEE standards.
Features:
 |
Cost-effective, flexible trace moisture analyzer with dew point and optiona l ppmv display |
 |
Dew point measurement range of -100°C to +20°C. with ±1°C accuracy |
 |
Interference-free PFM signal transmission
• Compact, 7 pitch (35 mm) module |
 |
Standardized, programmable probe output signal; with three alarms |
 |
ACKBUS compatible
|
原廠目錄下載
(PDF Format, 1 pages, 96 KB.)
HygroGuard
2650 Trace Moisture Analyzer
|
|
SPECIFICATIONS:
Probe
Sensor: Planar gold/aluminum
capacitance
Calibration Range: -80°C to +20°C
(-112°F to +68°F)
special calibration to -100°C (-148°F)
Recommended Recalibration Cycle: 6-12
months depending on the application
Accuracy: ± 2°C ( ± 3.6°F) in the
standard calibration range
Repeatability: Better than 1°C ( ±
1.8°F)
Max. Relative Humidity: 50% @ dew
points > 0°C (32°F)
Temperature Sensor: Zener device, range
-70°C to +70°C (-94°F to +158°F)
Signal Transmission: Frequency,
generated by probe electronics
Operating Pressure: 0
to 3000 psig (0 to 200 bar)
Recommended Flow Rate: 1 to 5 SCFH (in
a bypass mode)
Gas Flow Velocity: Static up to 165
ft/sec @ 14.7 psi, 16.5 ft/sec @ 150 psig,
1.65 ft/sec at 1500 psig with no particles in the stream; higher with
sintered end cap
Probe Tube: 1/2” diameter, 316
stainless steel
Mounting Adapter: 1/2” tube x 1/2”
MNPT, 316 SS compression fitting standard
Cable Connection: Rugged multi-pin
connector; screw terminal in explosion-
proof junction box with 1/2” FNPT conduit connection
Probe Cable: 4-conductor, AWG 22,
stranded, shielded to maintain EMI/RFI/ESD resistance, up 1000 ft., 5
ft. standard
Weight: .45 kg. (1 lb.)
2550 Analyzer Specifications
Electronics: Self-checking,
microprocessor-controlled, update measured
value every 2 sec.
EMI/RFI/ESD Protection: Exceeds IEC
801.2, ANSI C63, IEC 801.4, ANSI/IEEE
C62.41, and 801.5, ANSI/IEEE C62.41 standards
Accuracy: ± 0.2%
Moisture Units: Dew point in °C or °F
and ppmv, user programmable
Instrument Range: -130°C
to +20°C (-202°F to +68°F) dew point temperature, 0.001 to 9999 ppmv
Temperature Units: °C or °F, range
-20°C to +80°C (-4°F to +176°F)
Outputs: Isolated
output of 0/4-20 mA into 500 ohms max.,
programmable range corresponding to user-selected moisture unit; two
alarm relays, fail-safe mode, SPDT, dry contacts rated at 250 V AC, 2.5
A, P AC = 300 VA, cos >0.7 P DC = 100 W, 100 V DC; alarm set points,
output range, pressure constant
Cable Connection: Screw terminal (NEMA
4X includes 4 cable glands)
Program: Nonvolatile
memory
Data: EEPROM
User Interface: Three push-buttons to
select moisture or temp. reading,
moisture or temp. units (dew point, ppmv, °C, °F), alarm set points,
two temp. calibration points, output range, pressure constant
Display: 1/2” four-character LCD
Operating & Storage Temperature: -10°C
to +50°C (14°F to 122°F)
Power: 85 to 130 V AC, 187 to 276 V AC,
jumper selectable; 7 VA; 50/60 Hz; 24 V DC optional
Weight: 1.8 kg (4 lbs.)
Ordering Information Analyzer
2550
1 2 3 4
1 Certification
O Standard configuration (no approvals)
CG CSA General Purpose Approved
2 Analyzer Without Probe
1 No enclosure
4 Panel mount
5 NEMA 4X/CSA Type 3 with 4 cable glands (Fig. 3)
7 NEMA 3/4/7/9, Class I, Div. 1 & 2, Groups B, C, & D
8 Panel mount with door (Fig. 4)
3 Alarm
D 2 Alarm relays, 250 V AC, 2.5 A, both assigned to moisture
(programmable set points)
4 Power
0 115 V AC, 50/60 Hz
1 220 V AC, 50/60 Hz
2 24 V DC
Ordering Information Moisture
Probe DY – 5
1 2 25 3 5
1 Mounting Adapter
0 None
2 1/2” MNPT X 1/2” tube, 316 SS compression fitting
3 1/2” MNPT X 1/2” tube 316 SS compression fitting
4 1/2” X 16 “O” ring X 1/2” tube 316 SS compression fitting
6 SS – 8 - VCR( - 3 gland with female nut SS – 8 - VCR( - 1 welded
to the probe tube, includes gasket/retainer assay
2 Probe Structure
1 Replacement moisture probe for explosion-proof probes: 1/2” diameter
5” long 316 SS probe with 1/2” MNPT welded on for junction box
mounting
2 Moisture probe mounted to an explosion-proof junction box for
hazardous areas Class I, Div. 1 & 2, Group C & D, overall length
10” (Specify on order if for Group B)
7 Probe with 8-pin cable connector, overall length 7.8”
3 Protective Sensor Cap
D Standard end cap
S Sintered stainless, 100(m (use with high flow, particulates and
dryers)
5 Certification
0 None
C1 Designed for Class I, Div. 1, Gr. B, C, D when used with intrinsic
safety barriers
C2 Designed for use in Class I, Div. 2, Gr. B, C & D areas
CX Explosion-proof with flame arrestor-equipped flow cell for Class I,
Div. 1, Gr. C & D
Ordering Information Interconnecting
Cable
1 /2550 2
1 7 Multi-pin connector and tinned ends (use with probe DY-5_7 25)
2 Spade lugs and tinned ends (use with probe DY-5_2 25)
2 Cable Length
O 5 ft. long
L Other length (specify required length up to 1000 ft.)
|