水份濕度露點溫度檢測儀器 系列 Relative Humidity & Dew Point Measurement 
 
 

 

鏡面冷凝式露點感應器 / Chilled Mirror Dew Point Sensor 

管內嵌入式一段冷卻鏡面冷凝露點感應器

原廠目錄下載 (PDF Format, 1 pages, 75 KB.)
Chilled Mirror Sensor Model 1111H-GE

Chilled Mirror Sensor Model 1111H-GE

Description

The Model 1111H-GE is a duct mounted single stage dew point sensor with 45ºC (81ºF) of depression capability. It is equipped with a mounting flange, measures 1.5” in diameter and extends 9” into the duct. The 1111H-GE is a laboratory grade humidity transfer sensor suitable for use in industrial environments. The probe is equipped with a field replaceable PTFE membrane filter and fully recovers from condensation. It has been tested both in the lab and in the field to withstand the rigors of industrial and outdoor environments. As with all of General Eastern’s chilled mirror sensors, the 1111H-GE provides a direct primary measurement of dew point resulting in a humidity measurement that is certified as a NIST traceable primary reference standard.

Features and Benefits
項目符號 Primary, fundamental measurement of dew/frost point
項目符號 Hermetically-sealed, wire-wound 4-wire PRT
項目符號 Equipped with field replaceable PTFE Filter
項目符號 Suitable for use in industrial and outdoor environments
項目符號 Fully recovers from condensation
項目符號 Interchangeable with other General Eastern sensors without recalibration
項目符號 Functions with any General Eastern M series monitor 
項目符號 PACER Cycle for self-cleaning & rebalancing
項目符號 Field-replaceable mirror

嵌入式一段冷卻鏡面冷凝露點感應器

Model 1111H
The Model 1111H is a general purpose, 1-stage sensor with 45º C (81º F) of depression capability. Its 1" NPT mount allows it to be installed directly in line in a standard 1" pipe tee. Alternatively, it can be mounted in the 0111D pressure boss (200 psig max.), or
wall-mounted using the MB-11 adapter bracket.

Features:

項目符號

Primary, fundamental measurement of dew/frost point.

項目符號

Hermetically-sealed, wire-wound 4-wire PRT

項目符號

NIST traceability

項目符號

Interchangeable with other General Eastern sensors without recalibration

項目符號

Continuous, on-line measurement

項目符號

Infrared optics

項目符號

Field-replaceable mirror

項目符號

45ºC (81ºF) depression capability.
 

Accessories:

0111D, pressure boss

MB-11, wall mounting bracket
HEX-1, rhodium-plated mirror (standard)
HEX-2, solid platinum mirror
TW-1, mirror torque wrench
1123HK, sensor cable (10 ft (3 m)) standard, up to 300 ft. (91 m) available.

SPECIFICATIONS:

Performance

Sensing Element:  1/3 Class A DIN 43760 PRT, 100 ohm @ 0ºC (32ºF)
Dew/Frost Point Accuracy
- Standard: +/- 0.2ºC(+/- 0.36ºF)
- Optional: +/- 0.15ºC (+/- 0.27ºF)
Sensitivity: > 0.03ºC (0.05ºF)
Repeatability: +/- 0.05ºC (+/-0.09ºF)
Hysteresis: None
Cooling Stages: 1
Depression @ 25º, 1 atm, air: 45ºC(81ºF)

Typical Meas. Range @ 25ºC(77ºF): 1 atm, air:
Dew/Frost Point: -15º to +25ºC (+5º to +77ºF)
Relative Humidity: Equivalent to 6% to 100%
Required Sample Flow: 0.5 to 5.0 scfh (0.25 to 2.5L/m

Functional
Power: From electronics
Operating Ranges:
- Ambient Temperature: -15º to +80ºC(+5º to +176ºF)
- Pressure: -3 to 200 psig (0.8 to 15 bar)
Auxiliary Cooling: Not available

Physical:
Sensor Body Material: Epoxy-coated aluminum
Sensor Wetter Material: Aluminum, copper, Mylar, Teflon, rhodium or platinum
Weight: 1 lb. (0.5 kg)
Shipping Weight: 3 lbs. (1.4 kg.)

原廠目錄下載 (PDF Format, 1 pages, 37 KB.)
1-Stage Chilled Mirror Sensor, Model 1111H


連續流通式二段冷卻鏡面冷凝露點感應器

原廠目錄下載 (PDF Format, 2 pages, 245 KB.)
Chilled Mirror Sensor Model 1211HS - GE

Chilled Mirror Sensor Model 1211HS - GE

Description

Model 1211HS is a general purpose, 2-stage chilled mirror sensor with 65ºC (117ºF) of depression capability. Model 1211HS can be used as in the laboratory as well as industrial applications. The sensor is available with an auxiliary liquid cooling jacket to extend the range resulting in an economical standard for the validation of compressed and process air to -60°C frost point. The 1211HS functions with any General Eastern M or E series monitors to provide a time proven hygrometer system that provides primary humidity measurement. The hygrometers system is calibrated to NIST traceable reference standards and is suitable for use as humidity transfer standard. General Eastern’s manufacturing and quality system is certified to ISO-9001 standards. For more information on the theory and operation of General Eastern’s Chilled Mirror Hygrometers and Humidity Calibrators, contact us to obtain a catalog and application assistance.

Applications
項目符號 Laboratory Calibration Standards
項目符號 Compressed Air Dryers
項目符號 Emissions Testing
項目符號 Instrument Air Lines
項目符號 Heat Exchanger & Coil Calorimetry
項目符號 Environmental Chambers
項目符號 Heat Treat Furnaces

連續流通式二段冷卻鏡面冷凝露點感應器

Model D-2
The Model D-2 is a general purpose, 2-stage sensor with 65ºC (117ºF) of depression capability. It features wetted parts of stainless steel and glass, for durability in demanding industrial applications. The Model D-2 can be used as a benchtop sensor, mounted to a heat sink, or mounted to a cooling fan, if desired, for maximum operating range. Advanced features include field-replaceable optics and cooler assemblies, and auxiliary, visible light optics with viewing window for inspecting the mirror during operation.

Features:

項目符號

Primary, fundamental measurement of dew/frost point.

項目符號

Hermetically-sealed, wire-wound 4-wire PRT

項目符號

NIST traceability

項目符號

Interchangeable with other General Eastern sensors without recalibration

項目符號

Continuous, on-line measurement

項目符號

Infrared optics

項目符號

Auxiliary, visible light optics and viewing window for critical work

項目符號

Field-replaceable mirror, optics, and cooler assemblies

項目符號

65ºC (117ºF) depression capability

項目符號

Wetted parts of corrosion-resistant materials

項目符號

Flip-top lid with latch for easy access. 
 

原廠目錄下載 (PDF Format, 1 pages, 41 KB.)
2-Stage Chilled Mirror Sensor, Model D-2 

SPECIFICATIONS

Performance
Sensing Element:
 1/3 Class A DIN 43760 PRT, 100 ohm @ 0ºC (32º)
Dew/Frost Point Accuracy
-
Standard: +/- 0.2ºC(+/- 0.36ºF)
-
Optional: +/- 0.15ºC (+/- 0.27ºF)
Sensitivity: > 0.03ºC (0.05ºF)
Repeatability: +/- 0.05ºC (+/-0.09ºF)
Hysteresis: None
Cooling Stages: 2
Depression @ 25ºC, 1 atm,air: 65ºC (117ºF)

Typical Meas. Range @ 25ºC(77ºF): 1 atm, air:
Dew/Frost Point: -35º to +25ºC (-31º to +77ºF)
Relative Humidity: equivalent to 1.5% to 100%
Required Sample Flow: 0.5 to 5.0 scfh (0.25 to 2.5 L/min)

Functional
Power: From electronics

Operating Ranges:
Ambient Temperature: -25º to +85ºC (-13º to +185ºF), consult factory. for higher temp. requirements
Pressure: -14.7 to 185 psig (0 to11 bar)
Auxiliary Cooling: Not available

Physical:
Sensor Body Material: Stainless Steel
Sensor Wetter Material: 302, 316SS, Silicone O-ring, BK-7 glass, rhodium or platinum mirror
Weight: 1.8 kg. (4 lb.)
Shipping Weight: 3.2 kg. (7 lb.)

Accessories:
HEX-3, rhodium-plated mirror
HEX-4, solid platinum mirror
TW-1, mirror torque wrench
D-2K, sensor cable (10 ft (3 m) standard), up to 300 ft (91 m) available
OP-D2 optics module
CB-D2 cooler module
VB-D2 vapor barrier.


連續流通式四段冷卻鏡面冷凝露點感應器

Model 1311DR
The Model 1311DR is a 4-stage sensor designed to measure low frost points. It can be air-cooled giving 95ºC (171ºF) of depression capability, or liquid-cooled giving 105ºC (189ºF) of depression capability.

Features:

項目符號

Primary, fundamental measurement of dew/frost point.

項目符號

Hermetically-sealed, wire-wound 4-wire PRT

項目符號

NIST traceability

項目符號

Interchangeable with other General Eastern sensors without recalibration

項目符號

Continuous, on-line measurement

項目符號

Infrared optics

項目符號

Field-replaceable mirror

項目符號

95º C (171º F) depression capability; Air-cooled

項目符號

105º C (189º F) depression capability; Liquid-cooled.
 

Accessories:

HEX-1, rhodium plated mirror (standard)
HEX-2, solid platinum mirror
TW-1, mirror torque wrench
1123HK, sensor cable (10 ft (3 m) standard, up to 300 ft (91 m) available.

原廠目錄下載 (PDF Format, 1 pages, 36 KB.)
4-Stage Chilled Mirror Sensor, Model 1311DR

SPECIFICATIONS

Performance

Sensing Element: 1/3 Class A DIN 43760 PRT, 100 ohm @ 0ºC (32º)
Dew/Frost Point Accuracy:
-
Standard: +/- 0.2ºC(+/- 0.36ºF)
-
Optional: +/- 0.15ºC (+/- 0.27ºF)
Sensitivity: > 0.03ºC (0.05ºF)
Repeatability: +/- 0.05ºC (+/- 0.09ºF)
Hysteresis: None
Cooling Stages: 4
Depression @ 25ºC, 1 atm, air:
Air-cooled: 95ºC (171ºF)
Liquid-cooled: 105ºC (189ºF)
Typical Meas. Range, Air-cooled, 25ºC (77ºF),1 atm, air:
Dew/Frost Point: -65º to +25ºC (-85º to +77ºF)
Relative Humidity: equivalent to 0.03% to 100%
Liquid-cooled (15ºC (59ºF) cooling water, at 25ºC (77ºF), 1 atm air):
Dew/Frost Point:
-75º to +15ºC (-103º to +59ºF)
Relative Humidity: equivalent to 0.007% to 100%
Required Sample Flow: 0.5 to 5.0 scfh (0.25 to 2.5 L/min)

Functional
Power: 115/230 VAC, 50-60 Hz, 300 watts max.
Operating Ranges:
Ambient Temperature: 0º to +35ºC(+32º to +95ºF)
Pressure: 0 to 300 psig (1 to 22 bar)
Auxiliary Cooling: Water cooling (standard)

Physical:
Sensor Body Material:
Type 314 stainless steel
Sensor Wetter Material: 316SS, copper, mylar, silicone, rhodium or platinum mirror
Inlet/Outlet Fittings: 1/4" compression, 6 mm available
Weight: 16 kg. (34 lb.)
Shipping Weight: 18 kg. (40 lb.)


連續流通式五段冷卻鏡面冷凝露點感應器

Model 1311XR
The Model 1311XR is General Eastern's top of the line, 5-stage sensor designed to measure ultra-low frost points. It provides an impressive 112ºC (202ºF) of depression capability, enabling measurements in the <500 ppb range. The Model 1311XR, mated with the Model M3 monitor, provides a state of the art, ultra-low frost point measurement at an affordable price.

Features:

項目符號

Primary, fundamental measurement of dew/frost point.

項目符號

Hermetically-sealed, wire-wound 4-wire PRT

項目符號

NIST traceability

項目符號

Interchangeable with other General Eastern sensors without recalibration

項目符號

Continuous, on-line measurement

項目符號

Infrared optics

項目符號

Field-replaceable mirror

項目符號

112ºC (202ºF) depression capability

Accessories:

HEX-1, rhodium plated mirror (standard)
HEX-2, solid platinum mirror
TW-1, mirror torque wrench
1123XRK, sensor cable (10 ft (3 m) standard, up to 300 ft (91 m) available.

原廠目錄下載 (PDF Format, 1 pages, 38 KB.)
5-Stage Chilled Mirror Sensor, Model 1311XR

SPECIFICATIONS

Performance
Sensing Element: 1/3 Class A DIN 43760 PRT, 100 ohm @ 0ºC (32ºF) 
Dew/Frost Point Accuracy
-
Standard: +/- 0.15º C(+/- 0.27ºF)
-
Sensitivity: > 0.03º C(0.05ºF)
Repeatability: +/- 0.05ºC (+/- 0.09ºF)
Hysteresis: None
Cooling Stages: 5
Depression @ 25º, 1 atm,air: 112ºC (202ºF)

Typical Meas. Range with 15ºC (59º) cooling water, at 25ºC (77º), 1 atm, air:
Dew/Frost Point: -80º to +15ºC (-112º to +59ºF)
Relative Humidity: Equivalent to 0.0003% to 100%
Required Sample Flow: 0.5 to 5.0 scfh (0.25 to 2.5 L/min)

Functional
Power: 115/230 VAC, 50-60 Hz, 700 watts max.

Operating Ranges:
Ambient Temperature: 0º to +35ºC (+32º to +95ºF)
Pressure: 0 to 100 psig (1 to 8 bar)
Auxiliary Cooling: Water cooling (standard)


Physical:
Sensor Body Material: Stainless steel
Sensor Wetter Material: 316SS, copper, mylar, silicone, rhodium or platinum mirror
Inlet/Outlet Fittings: 1/4" compression, 6 mm available
Water Fitting: 3/8" compression
Weight: 23 kg. (50 lb.)
Shipping Weight: 26 kg. (58 lb.)


高溫環境用二段冷卻鏡面冷凝露點感應器

Model SIM-12H
The Model SIM-12H is a heated, 2-stage sensor with 65ºC (117ºF) of depression capability. It is designed for high dew point applications up to 85ºC (185ºF). For dew point measurements above ambient temperature, all components that come in contact with the sample gas must be heated to a temperature above the dew point, in order to prevent condensation and unreliable readings. For information on the Model HSS-12 heated sampling system, for use with the Model SIM-12H, refer to the Sampling Systems group on the Chilled Mirror Index page.

Features:

項目符號

Primary, fundamental measurement of dew/frost point.

項目符號

Hermetically-sealed, wire-wound 4-wire PRT

項目符號

NIST traceability

項目符號

Interchangeable with other General Eastern sensors without recalibration

項目符號

Continuous, on-line measurement

項目符號

Infrared optics

項目符號

Measures dew points above ambient, up to +85ºC (+185ºF)

項目符號

Field-replaceable mirror

項目符號

65ºC (117ºF) depression capability
 

原廠目錄下載 (PDF Format, 2 pages, 71 KB.)
Heated 2-Stage Chilled Mirror Sensor, Model SIM-12H

SPECIFICATIONS

Performance
Sensing Element:  1/3 Class A DIN 43760 PRT, 100 ohm @ 0º C (32º F)
Dew/Frost Point Accuracy
- Standard:
+/- 0.2ºC (+/- 0.36ºF)
- Optional:
+/- 0.15ºC (+/- 0.27ºF)
Sensitivity: > 0.03ºC (0.05ºF)
Repeatability: +/- 0.05ºC (+/- 0.09ºF)
Hysteresis: None
Cooling Stages: 2
Depression @ 25º, 1 atm, air): 65ºC (117ºF)

Typical Meas. Range - 75ºC (167ºF), 1 atm, air):
- Dew/Frost Point: -10º to +75ºC (+14º to +167º F)
- Relative Humidity: equivalent to 1% to 100%
Required Sample Flow: 0.5 to 5.0 scfh (0.25 to 2.5 L/min)

Functional
Power:
Sensor:
From electronics
Operating Ranges:
Ambient Temperature:
-15º to +50º C(+5º to +122ºF)
Control Temperature*:
+20º to +85ºC (+68º to +185ºF)
Pressure:
-3 to 50 psig (0.8 to 4.5 bar)
Auxiliary Cooling:
Not available

Physical:
Sensor Body Material:
Anodized Aluminum
Sensor Wetter Material: Aluminum, copper, plastic, mylar, rhodium or platinum mirror
Inlet/Outlet Fittings: 1/4" compression, 6 mm available
Weight: 3.2 kg (7 lb.)
Shipping Weight:
4.5 kg. (10 lb.)

Accessories:
HEX-1, rhodium plated mirror (standard)
HEX-2, solid platinum mirror
TW-1, mirror torque wrench
1123HK, sensor cable (10 ft (3 m) standard, up to 300 ft (91 m) available
HSS-12, heated sampling system

 

 

v  建成科學儀器股份有限公司     |      網 址: www.ssco.com.tw    |      服務電話:  (02) 29103818     |     傳 真:  (02) 29101316 v