Model GC-1 RH-Generator
Features:
 |
Built-in chilled mirror hygrometer for continuous
control |
 |
NIST and CETIAT traceability |
 |
Large 4 cubic
foot capacity chamber to accommodate RH
instruments |
 |
Continuous digital control |
 |
Prostep
software for fully automatic calibrations |
 |
Fluid-jacketed
constant temperature chamber |
 |
Fast, convenient, and cost
effective method to calibrate RH equipment |
 |
Clear Plexiglass chamber lid |
 |
Casters for convenience and
greater mobility |
 |
Available in Standard and Variable
Temperature models |
 |
Variable-temperature Model includes
chiller. |
原廠目錄下載
(PDF Format, 2 pages, 311
KB.)
Relative
Humidity Generator, Model GC-1
|
|
SPECIFICATIONS
Common to Both
Models
Voltage: 0100/115/230 VAC 50-60 Hz (specify) Volume of distilled water required: approx. 99 liters (approx. 26 gal.) Dry gas requirements: Frost point -40º
(-40º or lower), Clean oil-free gas regulated at 20 psig at
5-SCFH.
Dimensions: External: 1092 x 1626 x 686 mm (43"H x 64"W x 27.0"D) Test Chamber: 356 x 836 x 429 mm (14"H
x 32.9"W x 16.9"D) Test chamber
volume: 113.2 L (4 cu ft.) Cover material: Clear Plexiglass, can
be drilled and/or slotted by user for cable access Test chamber material: Anodized
Aluminum Programmability: Calibration profile may be programmed with Prostep
software Range: 10% to
90% RH Environment:
Should be temperature stable within +/- 1ºC between 20ºC and 30ºC (+/- 2ºF
between 68ºF and 86ºF) Time to set
point: approx. 60 minutes Accuracy: +/- 1% between 10 and 80% RH,
+/- 1.2% RH between 80 - 90% RH Stability: +/- 0.2% RH at 20 to 50% RH,
+/- 0.3% RH at 50 to 80% RH, +/- 0.5% RH at 10 to 20% and 80 to 90%
RH
Variable Temperature
Specifications
RH Accuracy: At chamber temps. of
30ºC to 60ºC (86ºF to140ºF) and 5º-20ºC (41º to 68ºF): +/- 1.5% RH between
20% - 80% RH, +/-2.0% RH between 10 and 20% and 80 to 90% RH. At chamber
temps. of 20º to 30ºC (68º to 86ºF): +/- 1.0% RH between 10% - 80%
RH,
+/-1.5% RH between 10 and 20% and 80 to 90% RH.
RH Stability: At chamber temps. of 20ºC
to 30ºC (68ºF to 86ºF): +/- 0.2% between 20% and 80% RH, +/- 0.4% between
10 and 20% and 80 to 90% RH. chamber temps. of 5ºC to 20ºC (41ºF to 68ºF)
and 30ºC to 60ºC (86ºF to 140ºF: +/- 0.3% between 20% and 80%
RH, +/- 0.5%
between 10 and 20% and 80 to 90% RH.
Temperature Stability: At chamber
temperature of 5º to 20ºC and 30ºC to 60ºC (41ºF to 68ºF and 86ºF to 140ºF):
+/- 0.3ºC (+/-0.6ºF) At chamber temps. of 20ºC to 30ºC (68ºF to 86ºF): +/-
0.1ºC (+/-0.2ºF)
|